Table 3 Range of motion (mean and SD) for goniometer and 3D motion capture measurement

From: Minimal detectable difference of the finger and wrist range of motion: comparison of goniometry and 3D motion analysis

  

Measurement day 1

 

Measurement day 2

 
  

Motion Capture

Goniometer

 

Motion Capture

Goniometer

P value

Joint

Direction

MeanromV1

SDromV1

MeanromG1

SDromG1

P value

MeanromV2

SDromV2

MeanromG2

SDromG2

[°]

[°]

[°]

[°]

[°]

[°]

[°]

[°]

Radioulnar

P/S

116

10

174

8

< 0.0001

114

10

170

8

< 0.0001

Wrist

F/E

150

9

141

14

0.570

150

10

139

12

0.499

Wrist

R/U

51

7

56

10

0.663

51

6

58

10

0.573

MCP2

F/E

109

10

114

12

0.787

109

9

116

13

0.663

MCP3

F/E

114

10

114

11

0.991

113

10

115

13

0.902

MCP4

F/E

113

11

115

10

0.883

110

10

119

14

0.644

MCP5

F/E

115

16

120

14

0.795

112

13

123

19

0.624

PIP2

F/E

118

6

109

6

0.275

118

6

113

7

0.617

PIP3

F/E

119

6

109

6

0.234

118

6

112

7

0.487

PIP4

F/E

122

7

109

6

0.186

121

8

112

7

0.411

PIP5

F/E

105

7

101

5

0.705

106

9

103

8

0.775

DIP2

F/E

80

10

84

9

0.751

83

10

84

9

0.899

DIP3

F/E

93

9

95

12

0.927

95

9

94

9

0.921

DIP4

F/E

84

13

87

12

0.882

85

13

89

12

0.833

DIP5

F/E

89

13

89

10

0.996

88

12

90

11

0.906

IP

F/E

93

14

103

23

0.720

92

13

104

21

0.643

MCP1

F/E

72

15

71

14

0.947

70

12

72

11

0.964

  1. Significant difference between motion capture and goniometer measurement