Skip to main content

Table 3 Parameters results

From: Comparison of three different internal fixation implants in treatment of femoral neck fracture—a finite element analysis

Parameters SCAP DHS+DS CCS
The maximum displacement of the femur (mm) 7.764 8.0087 8.1479
The maximum displacement of the internal fixation (mm) 7.2799 7.3649 7.9592
Maximum femur stress (MPa) 136.71 119.6 116.32
Internal fixation maximum stress (MPa) 395.92 462.29 363.43
The rotation angle (°) 1.29 1.88 2.35