Skip to main content

Table 1 Displacement between the cuneiform and the metatarsal bones before and after performing the test protocol II in millimetre

From: Biomechanics of common fixation devices for first tarsometatarsal joint fusion—a comparative study with synthetic bones

 

IOFix® mean (SD)

Double bridge plate® mean (SD)

Pedus L Plate® mean (SD)

Initial displacement

0.058 ± 0.005

0.051 ± 0.003

0.027 ± 0.005

Final displacement

0.076 ± 0.017

0.062 ± 0.002

0.030 ± 0.007