Skip to main content

Table 1 Mean reversible implant-bone micromotion amplitudes (μm) and standard deviations (SD) after 40,000 loading cycles at transducers S1–S6 (both n = 6)

From: Primary stability of a shoulderless Zweymüller hip stem: a comparative in vitro micromotion study

Transducer

 

CBH

CBH bp

p value

  

Mean

SD

Mean

SD

 

S1

μm

23.9

34.4

15.1

8.1

0.125

S2

μm

45.6

15.6

67.1

45.7

0.313

S3

μm

12.1

7.1

18.0

13.2

1.000

S4

μm

12.3

14.0

39.9

69.5

0.584

S5

μm

6.6

3.3

9.9

8.3

0.813

S6

μm

9.1

4.7

12.8

16.8

1.000