Skip to main content

Table 3 Average differences between projective angle measurements made using K-wires (intraoperatively), the BGS (intraoperatively), and postoperative CT reformats (in degrees)

From: Clinical evaluation of a biomechanical guidance system for periacetabular osteotomy

 

K-wires—intraoperative

Intraoperative–postoperative

K-wires—postoperative

Adduction

6.7 (−8.6–16.5)

3.7 (−0.3–9.2)

10.5 (−2.8–21.3)

Extension

4.3 (−2.7–24.5)

−0.1 (−8.1–11.2)

4.2 (−7.6–19.9)

Anteversion

−7.0 (−27.7 to −0.4)

−0.5 (−6.2–4.4)

−7.5 (−24.5–0.01)