From: Comparison of cement pressurisation in flanged and unflanged acetabular cups
 | Results | P values |
---|---|---|
Simplex | Â | Â |
Peak pressure | Charnley > Exeter unflanged | significant (p < 0.01) |
 | Charnley > Exeter flanged | significant (p < 0.01) |
 | Exeter unflanged > Exeter flanged | not significant |
Mean pressure | Charnley > Exeter unflanged | not significant |
 | Charnley > Exeter flanged | significant (p < 0.05) |
 | Exeter unflanged > Exeter flanged | not significant |
CMW | Â | Â |
Peak pressure | Charnley < Exeter unflanged | significant (p < 0.05) |
 | Charnley > Exeter flanged | significant (p < 0.01) |
 | Exeter unflanged > Exeter flanged | significant (p < 0.01) |
Mean pressure | Charnley < Exeter unflanged | not significant |
 | Charnley > Exeter flanged | not significant |
 | Exeter unflanged > Exeter flanged | not significant |