Skip to main content

Table 2 Summary of Results

From: Comparison of cement pressurisation in flanged and unflanged acetabular cups

 

Results

P values

Simplex

  

Peak pressure

Charnley > Exeter unflanged

significant (p < 0.01)

 

Charnley > Exeter flanged

significant (p < 0.01)

 

Exeter unflanged > Exeter flanged

not significant

Mean pressure

Charnley > Exeter unflanged

not significant

 

Charnley > Exeter flanged

significant (p < 0.05)

 

Exeter unflanged > Exeter flanged

not significant

CMW

  

Peak pressure

Charnley < Exeter unflanged

significant (p < 0.05)

 

Charnley > Exeter flanged

significant (p < 0.01)

 

Exeter unflanged > Exeter flanged

significant (p < 0.01)

Mean pressure

Charnley < Exeter unflanged

not significant

 

Charnley > Exeter flanged

not significant

 

Exeter unflanged > Exeter flanged

not significant