Skip to main content

Table 1 Three-dimensional hip contact force components [N] during normal walking, as applied to the finite-element-models for each of the four different implantation configurations.

From: Influence of prosthesis design and implantation technique on implant stresses after cementless revision THR

Implantation Configuration Hip Contact Force Component
  x y z
A: 34 mm offset, 4° anteversion -611 -73 -2539
B: 44 mm offset, 4° anteversion -659 -100 -2449
C: 34 mm offset, 14° anteversion -639 -4 -2679
D: 44 mm offset, 14° anteversion -694 -24 -2592
  1. Positive force components act medial (+x), anterior (+y), superior (+z). A, B, C, D represent the four implant configurations.