Skip to main content

Table 1 Loads applied to the hip system

From: The effect of abductor muscle and anterior-posterior hip contact load simulation on the in-vitro primary stability of a cementless hip stem

 

Loading step

Cycles

Hip contact force (xBW)

Fabd

   

Fcc

Fap(xBW)

 

Group 1 No abductor

1

1-1000

0.4 to 2.3

0

n/a

 

2

1001-2000

0.4 to 2.3

-0.1 to +0.3

n/a

 

3

2001-3000

0.4 to 2.3

-0.1 to +0.6

n/a

Group 2 Abductor

1

1-1000

0.4 to 2.3

0

1.1

 

2

1001-2000

0.4 to 2.3

-0.1 to +0.3

1.1

 

3

2001-3000

0.4 to 2.3

-0.1 to +0.6

1.1

  1. Fcc is in the caudal direction, and a positive Fap is in the posterior direction. The peak loads (maximum and minimum) were defined based on published data [15], and scaled for a 70 kg individual.