Table 1 Strains for hip implant alone (Construct A) for validating the FE model.

From: A biomechanical assessment of modular and monoblock revision hip implants using FE analysis and strain gage measurements

Side Location Modular (EXP) Modular (FE) Monoblock (EXP) Monoblock (FE)
   700 N 2000 N 700 N 2000 N 700 N 2000 N 700 N 2000 N
Medial 1 125 456 127 453 251 696 264 682
  2 144 523 157 524 259 789 249 756
  3 126 400 126 409 924 2966 482 1390
  4 190 543 200 570 894 2927 471 1319
Lateral 5 76 288 77 276 274 780 277 781
  6 80 274 82 268 188 677 172 673
  7 119 349 128 361 947 3077 392 1288
  8 290 767 298 776 821 2730 389 1221
  1. Medial strains are compressive. Lateral strains are tensile. EXP = experimental values. FE = finite element analysis values.