Skip to main content

Table 1 Loading profile applied to each liner specimen.

From: Experimental and analytical validation of a modular acetabular prosthesis in total hip arthroplasty

Loading case Speed (mm/min) Cycles Maximum Load (N)
1 5 20 250
2 5 20 350
3 5 20 450
4 5 20 550
5 5 20 650
6 5 20 750
7 5 20 850
8 5 20 950
9 5 20 1050
10 5 20 1150
11 5 20 1250
12 5 20 1350
13 5 20 1450
14 5 20 1550
15 5 20 1650
16 5 20 1750
17 5 20 1850
18 5 20 1950
19 5 20 2050
20 5 20 2150
21 5 20 2250
22 5 20 2350
23 5 20 2450
24 5 20 2550
25 10 25 2650
26 10 25 2750
27 10 25 2850
28 10 25 2950
29 10 25 3050
30 10 25 3150
31 10 25 3250
32 10 25 3350
33 10 25 3500
34 10 25 3750
35 10 25 4000
36 10 25 4250
37 10 25 4500
38 10 25 4750
39 10 25 5000
40 10 25 6000
41 0.5 2 10000
TOTAL   882  
  1. In each loading case the load range applied goes from 0 N to the maximum load specified for each loading case.