Skip to main content
Figure 6 | Journal of Orthopaedic Surgery and Research

Figure 6

From: Development of an in vitro three dimensional loading-measurement system for long bone fixation under multiple loading conditions: a technical description

Figure 6

Example of the 3D components of rotational displacement between ends of the 50 mm test segment versus applied torque (-50 to +50 Nm): A) intact control, B) osteotomy, and C) ostectomy. Note that the system resolution was capable of distinguishing total axial rotation of only 0.25, 0.45, and 2.0 degrees, respectively for these three models with notably less CrCa and LM bending rotation.